Autoclave – FLOOR / TC – DA TYPE
The steam sterilizer is a highly safe sterilizing product with stable function, suitable for major hospitals, clinics and biomechanical research laboratories etc., for the sterilization of instrument cloth, glassware and liquid etc.
The inner chamber is designed to be made #305 stainless steel, and the outer chamber and the door cover is to be made of #304 stainless steel, and the outer layer will be separated by thermal insulation wool for heat preservation. The outside is covered by rectangular stainless steel sheets which provides a more aesthetic appearance and heat preservation.
Stainless steel door with radial lever can effectively lock the door cover tightly and prevent steam leakage.
Silicone door gasket is high-temperature resistant, durable, and without deformation.
Indicator : displays the processing status of the sterilization process according to its sequence when the machine is operating.
Practical Sterilization Function :
1.Preset Vacuum function is able to shorten the sterilization time.
2.Vacuum drying function is able to remove unwanted moisture quickly, and to reach the Bowie Dick test criteria.
3.Cloth / apparatus / Liquid, Sterilization Functional Switch Mode can be selected to switch according to different kind of materials to be sterilized.
4.Auto Water Supply Device : Electrical control system is able to detect and sense the water volume automatically. Use equipment for soft water purification can prolong the life cycle of sterilizer.
Safety Devices :
1.Auto Power-Off Protection Device for water shortage, when water runs out, the electrical control system will automatically out off the heating power to prevent heating device from overheating, dry-out or even meltdown.
2.Pressure controller, the maximum pressure in the boiler is limited to 2.2 ±0.1kg/cm². When the pressure reaches the set value, the heating power will be automatically off to prevent the pressure from rising.
3.Safety valve, when the pressure reaches 2.8 kg/cm², it will start exhaust of steam and release the pressure from the chamber.
4.Pressure lock, when the pressure of inner chamber exceeds 0.2 kg/cm², the clutch will be locked automatically, and the door cover will not be able to be opened, in order to keep the operator safe.
Safety switch of Door Cover, when the door cover is opened the steam in the outer chamber is unable to enter the inner chamber. It cannot be operated normally unless the door cover is closed tightly.
ISO 9001 ISO13485 CNS9788 IEC61010-1 GMP